Power Mode | |
Max. Average Power | 200 W |
Max. Average Power (1) | 250 W |
Min. Power | 0.3 W |
Power Resolution | 10 mW |
Noise Equivalent Power (NEP) | 15 mW |
Response Time (0-90%) | 3 sec |
Power Calibration Uncertainty | ± 3% |
Power Linearity (2) | ± 1% |
Single Shot Energy Mode | |
Max. Energy (with 100 ms pulse) | 250 J |
Min. Energy | 1 J |
Energy Resolution | 10 mJ |
Energy Calibration Uncertainty | ± 5% |
Absorber Specs | |
Aperture | 60 mm |
Type | SHC |
Absorber Spectral Range | 0.19 - 11 µm |
Calibration Spectral Range | 0.2 - 1.1 µm, 10.6 µm |
Max Power Density (3) | 17 kW/cm² @200 W |
Max Energy Density (3) | 5ms pulse width: 115 J/cm² 10µs pulse width: 4 J/cm² 10ns pulse width: 1 J/cm² |
General Characteristics | |
Cooling | Forced Air with Fan (a) |
Weight | 1.2 Kg |
Dimension | 100 x 100 x 90 mm |
Cable lenght - connector | 1.5 m - DB15 2.5 m – USB (U option) 1.5 m - RS232 (R option) |
Stand and Post | Light Duty Stand |
Notes | |
(1). 2 minutes max (2). Detector centrally irradiated @50% of useful surface. (3). Damage thresholds also depend on power level. Please see damage graphs for more details. | (a). 12V DC Power Supply Included |
Optionen konfigurieren
Bitte lesen Sie die Geschäftsbedingungen.