Power Mode (1)
Requested pulse duration (1)100, 200, 500 ms
Max. Average Power40 kW (a)
Min. Power500 W (b)
Power Resolution10 W
Noise Equivalent Power (NEP)10 W (c)
Response Time4-10 sec
Power Calibration Uncertainty (1)± 5%
Power Linearity (2)± 2%
Repeatability± 1%
Absorber Specs
Apertureø 35 mm
TypeC
Absorber Spectral Range1050 - 1090 nm
Calibration Spectral Range1070 nm
Max Power Density (3)15 kW / cm2 @ 20 kW (d)
12 kW / cm2 @ 40 kW (e)
Max Energy Density (3)3000 J / cm2 (d)
General Characteristics
CoolingConvection
Weight2.55 kg
Dimensionø 115 mm x 85 mm
Cable length - connector2.5 m – DB15
Notes
(1) Laser pulse duration must be controlled with high precision: shorter or longer duration will directly affect measurement uncertainty. Recommended Laser rise time < 200 μs.
(2) Detector centrally irradiated, 50% of useful surface.
(3) Damage thresholds also depend on power level.
(a) pulse duration 100 ms (20 kW with 200 ms) (8 kW with 500 ms).
(b) pulse duration 500 ms (1 kW with 200 ms) (2 kW wih 500 ms).
(c) pulse duration 500 ms (25 W with 200 ms) (50W with 500 ms).
(d) 200 ms pulse duration.
(e) 100 ms pulse duration.
 
Opciones
Categoría: Ordering code: EXTREME-40
Leer los términos y condiciones

Descripción

Power Probe from 400W up to 40KW, DB15 connector for operation with PLUS2.