| Power Mode | |
| Max. Average Power | 200 W |
| Nominal Sensitivity | 0.22 mV/W |
| Min. Power | 0.2 mW |
| Power Resolution | 10 mW |
| Noise Equivalent Power (NEP) | 10 mW |
| Response Time (0-90%) | 2.5 sec (a) |
| Power Calibration Uncertainty | ± 3% |
| Power Linearity (1) | ± 1.5% |
| Absorber Specs | |
| Aperture | 30 mm |
| Type | HPB |
| Absorber Spectral Range | 0.19 - 11 µm |
| Calibration Spectral Range | 0.19 - 2.1 µm, 2.94µm, 9 - 11 µm |
| Max Power Density (2) | 7 kW/cm² @200 W |
| Max Energy Density (2) | 5ms pulse width: 36 J/cm² 10µs pulse width: 1.2 J/cm² 10ns pulse width: 0.3 J/cm² |
| General Characteristics | |
| Cooling | Water (b) |
| Weight | 0.25 kg |
| Dimension | 60 x 60 x 23 mm |
| Cable length - connector | 1.5 m – Bare wire (c) 2.5 m – USB (U option) 1.5 m - RS232 (R option) |
| Notes | |
| (1). Detector centrally irradiated @50% of useful surface. (2). Damage thresholds also depend on power level. Please see damage graphs for more details. | (a). Accelerated by Laserpoint electronics (b). Water 1.5 liter/min (@ 22°C) (c). Available, on request, with DB15 connector. |
Configurer les options