Power Mode
Max. Average Power600 W
Max. Average Power (1)800 W
Min. Power0.5 W
Power Resolution10 mW
Noise Equivalent Power (NEP)25 mW
Response Time (0-90%)2 sec
Power Calibration Uncertainty± 3%
Power Linearity (2)± 1.5%
Single Shot Energy Mode
Max. Energy (with 100 ms pulse)800 J
Min. Energy2 J
Energy Resolution10 mJ
Energy Calibration Uncertainty± 5%
Absorber Specs
Aperture30 mm
TypeSHC
Absorber Spectral Range0.19 - 11 µm
Calibration Spectral Range0.2 - 1.1 µm, 9 - 11 µm
Max Power Density (3)19 kW/cm² @500 W
Max Energy Density (3)5ms pulse width: 115 J/cm²
10µs pulse width: 4 J/cm²
10ns pulse width: 1 J/cm²
General Characteristics
CoolingWater (a)
Weight0.6 Kg
DimensionØ 90 x 33 mm
Cable length - connector1.5 m - DB15
2.5 m – USB (U option)
1.5 m - RS232 (R option)
Stand and PostLight Duty Stand
Notes
(1). 2 minutes max
(2). Detector centrally irradiated @50% of useful surface.
(3). Damage thresholds also depend on power level. Please see damage graphs for more details.
(a). Water 3 liter/min (@ 22°C); admissible rate of temperature variation < 1 °C/min
 
Configure options

Calibration Included (CAL >300W.)

Select Calibration Range

Additional Range

Category: Ordering code: W-600-D30-SHC
Read the terms and Conditions

Additional information

Sign up for our Newsletter and receive the latest news from LaserPoint




Choose the area of your interest:*

Subscribe the newsletter*